Base Plate Replacement

Tools Required

Torque screwdriver, TORX T20
Torque screwdriver, TORX T8
Torque screwdriver, TORX T10
Box torque wrench, 5/8 inch
Box torque wrench, 5.5 mm
Box torque wrench, 7 mm

Removal Procedure

Refer to Figure 6-10 for this procedure.
Step 1. Remove the Option Interface as described in "Option Interface Replacement" on page 99.
Step 2. Remove the 3/4 Rack Module Top as described in "3/4 Rack Module Top Replacement" on
page 90.
Step 3. Remove the A3 Analog Daughter Board as described in "A3 Analog Daughter Board
Replacement" on page 94.
Step 4. Remove the A2 Analog Board as described in "A2 Analog Board Replacement" on
page 96.
Step 5. Remove the two TORX T10 screws (item 1) fastening the Shield Angle.
Step 6. Remove the Shield Angle (item 2) from the Base Plate.
Step 7. Remove the PPMC PCA as described in "PPMC PCA Replacement" on page 101.
Step 8. Remove the A1 Mother Board as described in "A1 Mother Board Replacement" on
page 103.
Step 9. Remove the Power Supply Assembly as described in "Power Supply Assembly
Replacement" on page 105.
Step 10. Remove the seven TORX T20 screws (item 3) fastening the Base Plate.
Step 11. Remove the two TORX T20 screws (item 4) fastening the Base Plate and the ground
cables. Then, remove the ground cables (item 5).
Step 12. Remove the Rack Module Bottom from the Base Plate.
Step 13. Remove the Rear Panel as described from Step 9 to Step 11 in "Rear Panel Assembly
Replacement" on page 109.
Step 14. Remove the Base Plate from the Rear Panel.

Replacement Procedure

Step 1. Reverse the order of the removal procedure.
Chapter 6
Replacement Procedure
Base Plate Replacement
107