Required Equipment - Agilent Technologies E4980A Service Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

General Information

Required Equipment

Required Equipment
Table 1-1 lists the recommended equipment for performance test, adjustment and
troubleshooting for E4980A.
Table 1-1
Recommended Test Equipment
Equipment
Frequency Counter
Digital Multimeter
Standard Capacitors Set
Standard Capacitors Set
Four Terminal Pair
Standard Resistor Set
4TP Open Termination
4TP Short Termination
Test Leads (BNC), 1 m
Test Leads (BNC), 2 m
Test Leads (BNC), 4 m
DC Power Supply
Handler Simulator
Scanner Simulator
Bias Interface Simulator
Bias Interface Cable
Interface Box
Cables
Adapter
*1.P:Performance Test A:Adjustment T:Troubleshooting
16
Critical Specifications
Frequency Range: 1 MHz
Time Base Error < 50 ppm/year
No Substitute
No Substitute
No Substitute
No Substitute
No Substitute
No Substitute
No Substitute
No Substitute
No Substitute
+ 5V, 0.1A
No Substitute
No Substitute
No Substitute
GP-IO Cable
No Substitute
BNC(m)-BNC(m) Cable, 122 cm
BNC(f)-BNC(f) Adapter
Dual Banana-BNC(f) Adapter
Recommended Model
Agilent 53181A
Agilent 3458A
16380A
16380C with option
001
Agilent 42030A
Agilent 42090A
Agilent 42091A
Agilent 16048A
Agilent 16048D
Agilent 16048E
Agilent E3610A
p/n 04278-65001
p/n 04278-65301
p/n 42841-65001
p/n 42841-61640
p/n 04294-61002
p/n 8120-1840
p/n 1250-1830
p/n 1251-2277
Qty.
*1
Use
1
P
1
P,A
1
P,A
1
P
1
P,A
1
P,A
1
P,A
1
P,A
1
P,A
1
P,A
1
T
1
T
1
T
1
T
1
T
1
P,A
1
P,A
4
P,A
1
P,A
Chapter 1