A1 Mother Board Replacement; Tools Required; Removal Procedure; Replacement Procedure - Agilent Technologies E4980A Service Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

A1 Mother Board Replacement

Tools Required

Torque screwdriver, TORX T20
Torque screwdriver, TORX T8
Torque screwdriver, TORX T10
Box torque wrench, 5/8 inch

Removal Procedure

Refer to Figure 6-8 for this procedure.
Step 1. Remove the 3/4 rack module top as described in "3/4 Rack Module Top Replacement" on
page 90.
Step 2. Remove the PPMC PCA as described in "PPMC PCA Replacement" on page 101.
Step 3. Disconnect the Mother-Analog cable (item 1) from A1 Mother Board.
Step 4. Disconnect the front USB cable (item 2) from A1 Mother Board.
Step 5. Disconnect the cable power control (item 3) from A1 Mother Board.
Step 6. Disconnect the power supply cables (item 4) from A1 Mother Board.
NOTE
Pull out the power supply cables straight without touching the hook on the cable connector.
Step 7. Disconnect the LCD cable (item 5) from Front Panel Assembly.
Step 8. Disconnect the fan cable (item 6) from FAN Assembly.
Step 9. Remove the 5/8 inch nut and washer (item 7) fastening the BNC connector.
Step 10. Remove the five TORX T10 screws (item 8) from the A1 Mother Board.
Step 11. Remove the A1 Mother Board toward front side.

Replacement Procedure

Step 1. Reverse the order of the removal procedure.
NOTE
When you fasten the BNC connector with the 5/8 inch nut and washer, set box torque
wrench's torque to 17.4 in-lb.
When you fasten the A1 Mother Board with the five TORX T10 screws, set screwdriver's
torque to 8.7 in-lb for left-under screw of item 8 in the Figure 6-8, and set screwdriver's
torque to 4.35 in-lb for other screws.
When you connect the LCD Cable to the A1 Mother Board, make sure the LCD Cable is at
the bottom of the Analog Source Cable.
Chapter 6
Replacement Procedure
A1 Mother Board Replacement
103