Error Messages

Error number: 1103
Error Messages
An error message is displayed in the instrument status display area in the lower-left part of
the screen. Pressing the cursor arrow key on the front panel or executing the :DISP:CCL
command clears the error message. Moreover, about a specific error message, when a
sweep is started again, the display of an error message may disappear. Errors caused by
operation of a front panel key simply appear on the display; with a few exceptions, these
are not stored in the error queue.
The log of an error message, a maximum of 100 pieces are recorded, and it can go back and
check from the oldest error. Executing the :SYST:ERR? command checks error logs. This
operation can be performed only by "SCPI" command. It cannot be operated from the front
panel.
An error with a positive error number is one uniquely defined for this instrument. On the
other hand, an error with a negative error number is basically one defined for common
GPIB devices in IEEE488.2.
A
1103
A1 EEPROM write error
An error is generated while writing data to A1 EEPROM. When this error occurs, contact
Agilent Technology's Sales and Service Office or the company from which you bought the
device.
B
62
Bias off, Turn bias on
DC bias sweep is performed with the DC BIAS function OFF.
-160
Block data error
Block data is improper.
-168
Block data not allowed
Block data is not allowed.
C
-140
Character data error
Character data is improper.
-148
Character data not allowed
Character data not allowed for this operation.
-144
Character data too long
Character data is too long (maximum length is 12 characters).
-100
Command Error
Improper command.
41
Correction measurement aborted
142
Appendix D