Replacement Procedure

Display Interface Replacement

Display Interface Replacement

Tools Required

Removal Procedure

Refer to Figure 6-19 for this procedure.
Step 1. Remove the 3/4 Rack Module Top as described in "3/4 Rack Module Top Replacement" on
page 90.
Step 2. Remove the PCA E4980 USB as described in "PCA E4980 USB Replacement" on
page 117.
Step 3. Remove the A3 Analog Daughter Board as described in "A3 Analog Daughter Board
Replacement" on page 94.
Step 4. Remove the A2 Analog Board as described in "A2 Analog Board Replacement" on
page 96.
Step 5. Remove the Front Panel Assembly as described in "Front Panel Assembly Removal" on
page 98.
Step 6. Remove the Inverter as described in "Inverter Replacement" on page 120.
Step 7. Remove the Front Panel, the Key Pad, and the Key Flex Circuit as described in "Front
Panel, Key Pad, and Key Flex Circuit Replacement" on page 122.
Step 8. Remove the four TORX T6 screws (item 1) fastening the LCD Display.
Step 9. Remove the LCD Display (item 2) from the Display Support with the Display Interface.
Step 10. Carefully remove the Display Interface (item 3) from the Display Support (item 4).
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE
When you fasten the LCD Display with the four TORX T6 screws, set screwdriver's torque
to 3.05 in-lb.
When you replace the LCD Display with a new one, stick new gaskets (P/N : E4980-25003
 2, E4980-25004  2) to surroundings on the LCD Display.
124
Torque screwdriver, TORX T10
Torque screwdriver, TORX T20
Torque screwdriver, TORX T6
Box torque wrench, 7 mm
Box torque wrench, 5/8 inch