Error Messages
Error number: 45
S
Signal source overload
The signal source is overloaded. The measurement data is not guaranteed to be valid. The
measurement result data status of the measurement result output is set to 3.
When measuring Ls-Rdc/Lp-Rdc, set both RANGE and DCR RNG properly, or
"OVERLOAD" will be displayed.
Signal source overvoltage
The signal source voltage exceeds its limit. Lower the voltage setting of the signal source.
Appendix D