Power Requirement

Replacing Fuse

Replacing Fuse

Fuse Selection

Select proper fuse according to the Table C-1.
Table C-1
Use the fuse that meets the following specifications.
UL/CSA type, Slo-Blo, 5x20mm miniature fuse, 3A 250V (part number: 2110-1017)
Spare fuses are available from Agilent Technologies sales office. To check or replace the
fuse, disconnect the power cable and pull out the fuse holder (refer to Figure C-1) towards
you (refer to Figure C-2).
Figure C-1
Fuse holder and power cable socket
Figure C-2
Removing the fuse
138
Fuse Selection
Fuse Rating/Type
3A 250Vac
UL/CSA type
Time Delay

Setting up the fuse

Power cable receptacle
Fuse holder
Fuse Part Number
2110-1017
USE ONLY WITH A 250V FUSE