Calibrate:results:spc - Tektronix TBS2000 Series Programmer's Manual

Digital oscilloscope
Hide thumbs Also See for TBS2000 Series:
Table of Contents

Advertisement

Syntax

CALibrate:RESults:SPC?

Group
Syntax
Related commands
Returns
TBS2000 Series Programmer
CALibrate:RESults:FACtory?
Examples
CALibrate:RESults:FACtory? might return CALibrate:RESults:FACtory INIT
indicating the instrument has not be calibrated.
Returns the status of the SPC operation. This query does not initiate a SPC.
Query only.
Calibration and Diagnostic
CALibrate:RESults:SPC?
*CAL?
on page 53
INIT indicates that SPC has never successfully completed.
PASS indicates that the last SPC operation passed.
FAIL indicates that the last SPC operation failed.
RUNNING indicates that the SPC operation is running.
Examples
CALibrate:RESults:SPC? might return :CALibrate:RESults:SPC INIT indicating
SPC has not be run successfully.
C commands
59