M commands

MEASUrement:IMMed:UNIts?

Group
Syntax
Returns
Examples

MEASUrement:IMMed:VALue?

180
Returns the units for the immediate instrument measurement. Query only.
Measurement
MEASUrement:IMMed:UNIts?
<QString> where the string is one of the following: VOLTS, VOLTS
SQUARED, SEC, HERTZ, PERCENT, DIVS, SAMPLES, OHMS, AMPS,
WATTS, MINUTES, DEGREES, UNKNOWN, AMPS SQUARED, HOURS,
DAYS, DB, BYTES, INVERSE HERTZ, IRE, V OVER V, V OVER A, VOLTS
WATTS, V OVER W, VOLTS DB, V OVER DB, A OVER V, A OVER A,
AMPS WATTS, A OVER W, AMPS DB, A OVER DB, WATTS VOLTS, W
OVER V, WATTS AMPS, W OVER A, WATTS SQUARED, W OVER W,
WATTS DB, W OVER DB, DB VOLTS, DB OVER V, DB AMPS, DB OVER
A, DB WATTS, DB OVER W, DB SQUARED, DB OVER DB, VOLTS SEC,
AMPS SEC, WATTS SEC, V OVER S, A OVER S, W OVER S.
MEASUrement:IMMed:UNIts? might return :MEASUREMENT:IMMED:UNIts
"s", indicating that the unit for the immediate measurement is seconds.
Returns the value of the measurement specified by the
MEASUrement:IMMed:TYPe command. The measurement is immediately taken
on the source(s) specified by a MEASUrement:IMMed:SOUrce1 command.
Query only.
NOTE. A change to HORizontal:MAIn:SCALe or CH<x>:SCALe will not
necessarily have taken affect if immediately followed by this command.