Page 1
TBS2000B Series Oscilloscope Declassification *P077154000* 077-1540-00...
Page 3
TBS2000B Series Oscilloscope Declassification Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
Page 4
Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material.
Terms .............................. Clear and sanitize procedures Main board ............................USB flash drive ..........................Charges ............................TekSecure ............................Memory devices ..........................Terminology ........................... Clear NAND Flash data using TekSecure ..................Media and data export devices ....................... Change log TBS2000B Series Declassification...
Page 6
Table of Contents TBS2000B Series Declassification...
(USB ports, Ethernet, and eSATA). These instructions describe how to clear or sanitize the memory devices and disable the data output interfaces. The instructions also describe how to declassify an instrument that is not functioning. Products The following Tektronix products are covered by this document: ■ TBS2072B ■...
Terms The following terms may be used in this document: ■ Clear: This eradicates data on media/memory before reusing it in a secured area. All reusable memory is cleared to deny access to previously stored information by standard means of access. ■...
Main board Remove the Main board and return the product to Tektronix. A new main board will be installed, and the instrument will be repaired and adjusted as necessary. For removal instructions, refer to the TBS2000B Series Digital Storage Oscilloscope Service Manual , part number 077-1150-XX, available on the Tektronix Web site at www.tek.com/manuals.
It also allows the oscilloscope to be calibrated in a non-secure site and then used in a secure area without the need for recalibration. Memory devices The following tables list the volatile and nonvolatile memory devices in the instrument. TBS2000B Series Declassification...
Microproces User data, Directly Written by Main board Yes Remove Remove 256 MB (std) sor system user settings processor power from power from memory system instrument instrument for at least for at least 20 seconds 20 seconds TBS2000B Series Declassification...
Page 12
4 MB instrument operations applicable, applicable, calibration does not does not data, serial contain user contain user number, and data or data or Option key. settings. settings. Clearing Clearing would disable would disable instrument instrument functionality. functionality. TBS2000B Series Declassification...
3 minutes. Are you sure you want to erase?" 4. Press Yes. It will begin TekSecure, and after it finishes, it will reboot the instrument. Media and data export devices The following table lists the data export devices in the instrument. TBS2000B Series Declassification...
Page 14
USB device port The USB Device control and data using USBTMC on rear of the port cannot be transfer to a PC instrument disabled. LAN Ethernet Supports remote Directly Rear panel connector control and data transfer to a PC TBS2000B Series Declassification...