H commands
HORizontal[:MAIn]:SECdiv
Group
Syntax
Arguments
HORizontal:MAIn:UNIts[:STRing]
Group
Syntax
156
Sets the time per division for the main time base. This command is identical to
the HORizontal:MAIn:SCAle command. It is provided to maintain program
compatibility with some older models of Tektronix s.
Horizontal
HORizontal: SECdiv <NR3>
HORizontal: SECdiv ?
HORizontal:MAIn:SECdiv <NR3>
HORizontal:MAIn:SECdiv?
<NR3> specifies the range from 1(or 2) ns to 100 s, depending on the model.
Examples
HORIZONTAL: SECdiv 2E-6 sets the main scale to 2 μs per division.
HORIZONTAL: SECdiv ? might return :HORIZONTAL:MAIN: SECdiv
2.0000E-06 indicating that the main scale is currently set to 2 μs per division.
Returns the current horizontal unit "s". Query only.
Horizontal
HORizontal:MAIn:UNIts?
HORizontal:MAIn:UNIts:STRing?
Examples
HORizontal:MAIn:UNIts? might return HORizontal:MAIn:UNIts SECONDS.
TBS2000 Series Programmer