T commands
Arguments
OCCURS specifies a trigger event if a runt of any detectable width occurs.
LESSthan sets the instrument to trigger if a runt pulse is detected with a width
less than the time set by the TRIGger:A:RUNT:WIDth command.
MOREthan sets the instrument to trigger if a runt pulse is detected with a width
more than the time set by the TRIGger:A:RUNT:WIDth command.
EQUal sets the instrument to trigger if a runt pulse is detected with a width equal
to the time period specified in TRIGger:A:RUNT:WIDth within a ±5% tolerance.
UNEQual sets the instrument to trigger if a runt pulse is detected with a width
greater than or less than (but not equal to) the time period specified in
TRIGger:A:RUNT:WIDth within a ±5% tolerance.
Examples
TRIGGER:A:RUNT:WHEN THAN sets the runt trigger to occur when the
instrument detects a runt in a pulse wider than the specified width.
TRIGGER:A:RUNT:WHEN? might return :TRIGGER:A:PULSE:RUNT:WHEN
OCCURS indicating that a runt trigger will occur if the instrument detects a runt
of any detectable width.
TRIGger:A:RUNT:WIDth
Sets or queries the width for a runt trigger.
Group
Trigger
Syntax
TRIGger:A:RUNT:WIDth <NR3>
TRIGger:A:RUNT:WIDth?
Related commands
TRIGger:A:RUNT:WHEn
on page 238
TBS2000 Series Programmer
239
Need help?
Do you have a question about the TBS2000 Series and is the answer not in the manual?