5 Prepare the system for a run
5.5 Set pressure alarms
Step
1
2
3
4
118
Action
In the System Control module, on the Manual menu, click Execute Manual
Instructions .
Result: The Manual instructions dialog box opens.
In the Instructions box, select Alarms:Alarm pre column pressure.
Click Enabled in the Mode field.
Enter the high pressure limit in the High alarm box.
•
Click Execute.
•
ÄKTA avant Operating Instructions 29101556 AB