Table D-1. Physical Block Parameters
Index Parameter name
20
DESCRIPTOR
21
DEVICE_MESSAGE
22
DEVICE_INSTAL_DATE
23
LOCAL_OP_ENA
24
IDENT_NUMBER_SELECTOR
25
HW_WRITE_PROTECTION
26
FEATURE
27
COND_STATUS_DIAG
33
FINAL_ASSEMBLY_NUM
34
DOWNLOAD_MODE
35
PASSCODE_LOI
36
LOI_DISPLAY_SELECTION
37
LOI_BUTTON_STATE
38
VENDOR_IDENT_NUMBER
39
LOI_PRESENT
40
HW_SIMULATE_PROTECTION
1.
If blank, parameter is not applicable to LOI.
Table D-2. Transducer Block Parameters
Index Parameter name
1
ST_REV
2
TAG_DESC
3
STRATEGY
PROFIBUS® PA Block Information
PROFIBUS® PA Block Information
LOI
DTM
name
location
Descriptor
N/A
Message
N/A
Installation
N/A
Date
LOI Enable
N/A
Ident Number
IDENT
Selector
HW Write
N/A
Protection
Optional
Device
N/A
Features
N/A
N/A
Final Assembly
N/A
Number
Factory
N/A
Upgrade
Password
PSSWD
Display
DISP
Selection
Button State
N/A
Vendor Ident
IDENT
Number
LOI Present
N/A
HW Simulation
N/A
Protection
LOI
DTM name
location
Static Revision
N/A
No.
Tag
N/A
Strategy
N/A
Definition
(1)
User-definable text to describe the
device.
User-definable message to the device or
application in plant.
Date of installation of the device.
Disable/enable the optional LOI
Specifies the cyclic behavior of a device
which is described in the corresponding
GSD file
Status of the security jumper
Indicates optional features implemented
in the device
Indicates the mode of a device that can
be configured for status and diagnostic
behavior
The same final assembly number placed
on the neck label
Puts the device into a manufacturer
mode for upgrading the device
Password for the LOI
Indicates process variables shown on the
local display
Status of the optional LOI buttons
0x3333
Parameter written during manufacturing
to indicate if an optional LOI is present
Status of hardware simulation jumper
Definition
(1)
The revision level of the static data
associated with block; the revision value
will be incremented each time a static
parameter value in the block is changed.
The user description of the intended
block application.
Grouping of function blocks.
November 2016