EX10xxA/RX10xx calibration is performed using the SFP web interface. When accessing an action
or entry page for the first time during a session, an Error page will appear. To clear this error, click
on the login link. The Login page will appear. In the Password field, enter the configured
instrument password. The default factory-programmed password is ex1048. For additional
information, refer to the General Web Page Operation in Section 4.
To reset the password to factory default, replace "index.cgi?" with "change_password.cgi?" in the
NOTE
browser's address bar.
T
C
IME
ONFIGURATION
When calibration is performed, the time and date of calibration is recorded, making it important to
ensure that the EX10xxA/RX10xx's time is configured properly. To determine the time, click on
the Time Configuration link in the command menu. If the time is displayed incorrectly, use the
following steps to configure the time from the Time Configuration web page.
1) Select the correct Time Zone from the drop menu.
306
F
10-3: A
IGURE
CCESSING THE
F
10-4: L
IGURE
OGIN
EX10
A V
ACE
XX
IA
E
P
RROR
AGE
EX10xxA Calibration