Vtex10Xxa_Set_Dio_Limit_Event_Latch - VTI Instruments EX1000A User Manual

Hide thumbs Also See for EX1000A:
Table of Contents

Advertisement

VTI Instruments Corp.
F
P
UNCTION
ROTOTYPE
ViStatus vtex10xxA_set_dio_limit_event_latch(ViSession vi, ViInt16 dio_channel, ViBoolean latch);
F
P
UNCTION
ARAMETERS
vi = contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
dio_channel = the channel of the digital I/O port to be affected. Valid input values: 0 through 7.
latch = a Boolean value indicating whether to operate in latch mode. If this parameter is set to "1", the specified
DIO channel will operate in latch mode. If set to "0", the specified DIO latch mode is not enabled.
D
I
R
V
ATA
TEM
ESET
ALUE
latch = 0 (for all DIO channels)
D
ESCRIPTION
This function enables or disables latch operation of a DIO channel linked as a limit event. The nominal operation of
a linked DIO channel is to reflect the latest limit evaluation. That is, it will be updated with every scan. In latch
mode, a transition out of the cleared state would remain, regardless of future limit evaluations, until it is cleared at
the beginning of a new acquisition.
E
XAMPLE
236

vtex10xxA_set_dio_limit_event_latch