Vtex10Xxa_Self_Cal_Clear_Stored - VTI Instruments EX1000A User Manual

Hide thumbs Also See for EX1000A:
Table of Contents

Advertisement

F
P
UNCTION
ROTOTYPE
vtex1048_self_cal_clear_stored(ViSession vi);
F
P
UNCTION
ARAMETERS
vi = contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
D
I
R
V
ATA
TEM
ESET
ALUE
Not applicable to this function.
D
ESCRIPTION
This function clears self-calibration data from nonvolatile memory. This operation does not clear the current self-
calibration data, only that in non-volatile memory. If this function is called and no non-volatile self-calibration data
is present, an error is generated.
E
XAMPLE
EX10xxA Function Calls

vtex10xxA_self_cal_clear_stored

www.vtiinstruments.com
213