MP85A process controller
y
Fig. 6.11: Tolerance band
Limit monitoring
From firmware version 2.22 and higher, the limit values can also be included
in the total result for process monitoring. This allows, e.g. minimum and
maximum values to be monitored at the start and/or end of a process. This
option is particularly useful in the evaluation methods "Envelope curve" and
"Tolerance band": The following diagram monitors whether the start and end
of the measurement curve are lying in the yellow marked areas.
y
LV2
LV1
Fig. 6.12: Limit value monitoring
TIP:
The online help of the PME Assistant and additional software modules contain
an accurate description of all evaluation criteria with examples, diagrams and
Tips and Tricks. Start the PME Assistant and call help with F1.
A2392−5.0 en
LV1
x
x
LV2