Step 1) "Fresh Air" Step - LANDTEC GEM5000 Operating Manual

Gas analyzer
Hide thumbs Also See for GEM5000:
Table of Contents

Advertisement

5K-MNL-GEM5000

Step 1) "Fresh Air" step.

Highlight the mix labeled "Fresh Air" and press the enter key
will appear. The pump will automattically turn on and you will see the values in the Current
column changes as the gas stablizes.
Several symbols are used on the Gas Check screens to indicate the status of the Gas Check.
Status
Meaning
Symbol
The Gas channel is disabled or cannot be zeroed
The Gas channel is out of tolerance and recommended to span the channel.
The Gas channel is within specification
The Gas channel is out of range and has failed the check. Also the value
is consider out of limits and will not allow for calibration of the channel to
occur.
Once the numbers settle down a bit, choose "Check" to begin the actual check process.
You will see the countdown timer while the gas is being checked. Once complete a screen
will appear indicating the status of the various gas channels.
At this point, you are prompted to either Retry the check, Calibrate the instrument or Record
the check event.