5K-MNL-GEM5000
Note:
The list of special action options displayed on the special action menu is
dependent upon device type, mode, and step in the reading sequence. Site
& ID Questions will only be available if established in LSGAM software.
Refer to 5.7 Create a Project / Select a Project.
The following actions may be available:
Action
Restart Process
Site Questions
ID Questions
System Pressure
Enter Temperature
Simple Gas Reading
Filter H
S
2
2)
Select the soft-key 'Exit' to exit this menu and return to the 'Main Gas Read
Screen'.
Page 80
Figure 28
Special Actions
Function
This action enables the operator to restart the current process
again from the beginning. The operator will return the Main Gas
Read Screen.
This action enables the operator to update site questions prior to
taking a reading.
This action enables the operator to update ID questions specific
to sample points prior to taking a reading.
GEM mode – This action enables the operator to measure
available system vacuum before the well's valve when taking a
reading.
5.7.3.
This action enables the operator to enter the temperature
manually if a temperature probe is not installed.
The Simple Gas Reading or "Pump & Read" is a Gas
Composition Only reading.
connect hose, 2) pull sample, 3) obtain gas composition only, 4)
store reading.
prior to starting a reading or by simply turning on the pump
rather than pressing "Start".
This function can allow for better CO readings by removing H
from the gas sample. Refer to 8.8 How to use an H
Copyright LANDTEC North America, Inc.
GEM5000 Gas Analyzer
Refer to ACCU-FLO well device diagram in Section
The reading workflow is simply 1)
This can be invoked by the special action menu
S
2
S filter.
2
V1.13