LANDTEC GEM5000 Operating Manual page 86

Gas analyzer
Hide thumbs Also See for GEM5000:
Table of Contents

Advertisement

GEM5000 gas analyzer
Fitting a pitot tube to a sample point when using an H
2)
Make sure the water trap filter is fitted as close as possible to the pitot tube. This
will help protect and trap moisture before it reaches the H
the water trap filter is clean and dry.
3)
Attach the sample tubing into the analyzer.
4)
To take flow readings using a pitot tube, please follow the instructions displayed
on the front of the instrument. The user will be prompted to complete each step
and should follow the instructions on screen or those initially provided with the
Pitot Tube.
Note:
It is important to center the pitot tube within the pipe and ensure it is
parallel with the pipe.
LANDTEC
5K-MNL-GEM5000