GEM5000 Gas Analyzer
5K-MNL-GEM5000
Initially you may see many red X in the gas screen because the current value varies widely from the
expected target. At this point, connect the 50%CH
/ 35%CO
bottle to the white port on the top of
4
2
the analyzer and
begin the flow of
gas.
Once
gas
flows, the values
will
probably
change.
Now the gases for CH
, CO
, & O
are much closer and are within a range that is able to be
4
2
2
calibrated.
Press Check to begin the process of checking the gas. The countdown will begin.
Once the timer countdown is complete a summary of the calibration check is displayed.