OMGEM5KN4.7.1
9.7.1
Gas Check in fresh air
Switch on the
instrument and
START
complete the
instrument warm up.
Let the pump run for 90 seconds
and select soft key 'check'
Message will display
'Please wait for timer
to reach zero'
Was the
calibration
successful?
Unsuccessful
Please see the section
9.11 – Calibration
troubleshooting
Page 60 of 70
Main Gas Read
Screen select the
'Menu' key
When complete a
message will
display 'Check
complete'
Read status field
on the right hand
side of the screen
Retry
Calibrate
Calibrate
Record
Successful
Result saved
Select
soft-key
'Exit'
END
GEM5000 Gas Analyzer
Select key 2
'Gas Check'
Select 'Fresh Air' by using
the enter key
Record
Select key 1
'Gas check'
Disconnect all
tubing
This feature records
that a gas check has
been completed but
does not calibrate
the analyser
Select
soft-key
'Exit'