Step 3) Spanning Oxygen - LANDTEC GEM5000 Operating Manual

Gas analyzer
Hide thumbs Also See for GEM5000:
Table of Contents

Advertisement

5K-MNL-GEM5000
The results will be saved. The calibration gas bottle can be disconnected.
To recap:
Step 1 Spanned CH
Step 2 Spanned CH
Step 3 will Span will span O

Step 3) Spanning Oxygen

To span the Oxygen channel to ambient air choose the gas mix "Oxygen High".
The "Oxygen High" gas mix is defined for just Oxygen and no other gases are defined. This
allows for oxygen to be spanned without spanning CH4, CO, H2S, or H2 to Zero. Press
Check to begin checking the O2. As the gas is displaced from the previous Span in the
instrument the current value will rise.
Page 102
to Zero.
4
to 50% & CO
to 35%. Spanned O
4
2
to a non-zero span-target.
2
Copyright LANDTEC North America, Inc.
GEM5000 Gas Analyzer
to Zero.
2
V1.13