Step 2) Span Methane, Carbon Dioxide, & Zero Oxygen - LANDTEC GEM5000 Operating Manual

Gas analyzer
Hide thumbs Also See for GEM5000:
Table of Contents

Advertisement

GEM5000 gas analyzer
Press Calibrate to:
Span the Methane to a Target value of Zero
Span the Oxygen to 20.9
Span CO
This concludes the first step of the calibration. Press Exit to return to the Gas Mix screen.
Now you will span CH
, CO
4
Step 2) Span Methane, Carbon Dioxide, & Zero Oxygen
Use the up
& down
until the mix "CH4 High" is highlighted green. Then press the enter
LANDTEC
, H
S, and H
Channels to a Target of Zero.
2
2
2
and span O
to zero.
2
2
arrow keys on the instrument to navigate through the gas mixes
5K-MNL-GEM5000
key.
Page 99