5K-MNL-GEM5000
3) Manual adjustment of the flow fail is available via this option and can be carried out
with use of 'Key 4 – Scroll left' ( ) less sensitive and 'Key 6 – Scroll right' (>) more
sensitive.
4) Select the soft-key 'Save' to store the setting or select soft-key 'Exit' to exit the
screen without saving the change.
5) The operator will return to the 'Operation settings' menu.
Note:
the life of the pump. The default setting displays the bar in the
center. BEFORE altering this setting, please contact Technical Support at +1 (909)
783-3636 or
7.3.7
display the 'Device Settings' menu. Press the soft key to display 'User Settings'.
2) Select 'Key 6 – Technician login' and the following screen is displayed:
Page 76
Changing the flow fail setting to a less sensitive setting may decrease
email technical@landtecna.com
Technician login
This option enables the operator to select from a pre-defined technicians or
to add a Technician ID. All subsequent readings will be tagged with this
Technician Login ID.
1) Select the 'Menu' button on the front of the analyzer to
Figure 27
Copyright LANDTEC North America, Inc.
GEM5000 Gas Analyzer
—
Figure 26
Adjust Flow Fail
—
Technician Login
V1.13