Calibration Set-Up; Calibration Equipment - LANDTEC GEM5000 Operating Manual

Gas analyzer
Hide thumbs Also See for GEM5000:
Table of Contents

Advertisement

GEM5000 gas analyzer
Warning

9.4 Calibration set-up

Warning
The regulator supplied with the calibration kit has been configured to deliver a fixed flow.
As the regulator's flow is factory set, it only requires a few turns to open, no adjustment
is necessary.
Warning

9.5 Calibration equipment

The following diagram displays the regulator, tubing and carrying case used for user
calibration:
LANDTEC
5K-MNL-GEM5000
Calibration gases can be dangerous.
For each gas used the appropriate material safety data sheet
must be read and understood before proceeding.
Do NOT attach the gas supply to the gas analyzer before
putting the analyzer into the 'Gas Check' screen. Select
'Check Spans' from the 'Operation Settings' menu.
Exhaust port
When the gas analyzer is being calibrated, there are two
possible exits for the gas, via the usual manner out of the
exhaust (yellow) port of the analyzer or through the optional
pressure relief valve. In cases of over-pressurization the port
on the pressure relief valve will open.
It is recommended that tubing is attached to the instrument
during a calibration or gas check. The exhaust tubing must
emerge in a well-ventilated area. Ensure there are no leaks
in the tubing and connections.
The calibration of the gas analyzer should be carried out in
a safe area with all necessary precautions taken when using
potentially dangerous, explosive or toxic gases.
5K-MNL-GEM5000
GEM5000 Gas Analyzer
Page 95