Sirona Dental Systems GmbH
Operating Instructions SIROLaser Advance
2 .
2 .
1.
62 60 504 D3545
D3545.201.01.08.02 08.2012
7. Insert the handpiece body back into the handpiece sleeve making
sure that the optical fiber slides easily through the single-use tip. Also
check if the snap tabs have clicked into their rests.
8. Trim the end of the optical fiber with the fibercutter until the light of the
aiming beam projects an uniform circular pattern, see chapter
„Adjusting the optical fiber with the fiber cutter" [➙ 31]. To do this, aim
the optical fiber vertically at a white background located approx. 5 cm
away. If the probe projects no pattern at all or only an uneven pattern,
cut off one to two millimeters. Always make the notch perpendicular
to the optical fiber.
WARNING
If the optical fiber does not protrude at least 1 cm out of the single-use
tip, there is a risk that the single-use tip will heat up.

Adjusting the optical fiber with the fiber cutter

4.4.2.3.4
1. Place the optical fiber in the fiber cutter at the notched mark.
2. Press the fiber cutter together.
3. Bend the optical fiber at the notched location.
 The optical fiber breaks at the notched location with a smooth,
perpendicular fracture surface.
Check to see if the light of the aiming beam projects a uniform circular
pattern. To do this, aim the optical fiber vertically at a white background
located approx. 5 cm away. If the probe projects no pattern at all or only
an uneven pattern, cut off another one to two millimeters.
NOTICE
3 .
Press firmly but do not squeeze the optical fiber. You just need a little
notch to receive a perfect result when breaking the fiber at the notch.
4.4.3
Install wireless foot control – optional
The SIROLaser Advance can be operated using the finger switch (which
is integrated in the handpiece) as well as by using the optional wireless
foot control.
NOTICE
The foot switch has an IPX5 degree of protection. Therefore this foot
switch must not be used in hospital operating rooms.
Technical data of the wireless foot control, see chapter Technical Data,
"Wireless foot control".
The wireless foot control must be assigned to the SIROLaser Advance via
a registration. This prevents malfunctions caused by neighboring wireless
controls.
Initial start-up – procedure for proper assembly
Installation
31