 Maintenance and service
Replacement of parts subject to wear and tear
70
● Fiber cutter (change after 400 treatments/sterilizations or every two
years)
● Rechargeable battery (change after 1000 charging cycles or every
two years)
● Batteries in wireless foot control (change after 1 year)
For further informations, see chapter "Cleaning, disinfection and
sterilization" [ ➙ 62].
Only original spare parts, Sirona
CAUTION
Only use parts from Sirona Dental Systems, see "Spare parts" [ ➙ 24].