 Cleaning, disinfection and sterilization

Cleaning the control unit

64
NOTICE
Remove any possible water residues.
The parts must be sterilized only in an autoclave with saturated water
vapor at minimum sterilization values of 134°C (273.2°F), 3 min. holding
time and 2.04 bar (29,59 psi) overpressure.
Steam sterilizers are approved for sterilization that fulfill the
requirements of EN 13060 class B or validated steam sterilizer (EN
13060 class S) which are employing three, separate initial vacuum air-
purges being suitable for the sterilization of dental handpieces. For
example SIRONA DAC PROFESSIONAL.
Please use the transport and sterilization tube for sterilizing the optical
fiber.
CAUTION
Always protect the optical fiber with the transportation and sterilization
tube as well as the sterilization cap.
The optical fiber may be damaged if it is seriously bent or improperly
routed inside the handpiece. This may constitute a health hazard for
patients, dentists and dental assistants. The minimum bending radius of
the optical fiber is 4.5 cm (diameter: 9 cm). Be careful not to pinch or tear
the optical fiber when inserting or cleaning it.
7.4
Cleaning the control unit
Use a dry, soft cloth to remove dust from the SIROLaser Advance. More
stubborn spots can be removed with a damp cloth.
NOTICE
Please proceed careful not to scratch and damage the foil on the touch
screen.
You can wipe-disinfect the SIROLaser Advance using any of the products
that are commonly used to disinfect medical electrical equipment, e.g.
MinuteWipes, Caviwipes.
CAUTION
Spray disinfection may allow liquids to penetrate into the SIROLaser
Advance!
The SIROLaser Advance may be disinfected only by wiping it. Do never
spraydisinfect the SIROLaser Advance.
Observe the instructions provided by the manufacturers of these
disinfectants.
MinuteWipes Fa. Alpro.
In USA: Caviwipes
Operating Instructions SIROLaser Advance
.
Sirona Dental Systems GmbH
62 60 504 D3545
D3545.201.01.08.02 08.2012