Battery calibration

5.5.4.12
For perfect battery performance a battery calibration must be proceeded
in any case of having removed and again mounted or completely
replaced the battery pack, see chapter „Replacing the rechargeable
battery of the control unit" [ ➙ 68].
1. Switch on the laser without having connected the power supply.
2. Choose "Battery calibration" in the set-up menu.
 The following message will appear. "Unplug the laser and press
OK for battery calibration. For further steps refer to the user
manual."
3. Press 'OK'.
 The battery will now be discharged automatically until the device
switches off due to lack of power. You will also be informed that
"Battery calibration may take several hours."
4. When the device has been switched off automatically plug in the
power supply and charge the battery for at least 2 hours (best over
night).
 Now the battery is calibrated.