Usb Port; Time To Service; Calibration Check; Sirona Dental Systems Gmbh - Sirona SiroLaser advance Operating Instructions Manual

Hide thumbs Also See for SiroLaser advance:
Table of Contents

Advertisement

Sirona Dental Systems GmbH

Operating Instructions SIROLaser Advance
empty
empty
empty
62 60 504 D3545
D3545.201.01.08.02 08.2012
NOTICE
If the finger switch is marked 'cross' for defective/missing please check
if the handpiece sleeve is properly connected to the handpiece body. If
the problem remains please check the cable connection to the
SIROLaser Advance control unit, see chapter "Troubleshooting of
simple defects". If the finger switch remains marked 'cross' for defective/
missing please contact your local dental depot or an authorized
Customer Service Department for technical support.
In general: If any switch is defective laser will be blocked.

USB port

To make sure that the USB port is available it is checked within the self
test.
The USB port can be 'tick' for checked or 'cross' for defective.
With history file
NOTICE
If the USB port is marked 'cross' for defective please contact your local
dental depot or an authorized Customer Service Department for
technical support.
The SIROLaser Advance will remain functional but the download of the
history file is not possible.

Calibration check

Powermeter, Sirona
Sirona Dental Systems recommends a calibration check with external
powermeter to be performed every twelve months, see chapter "Country
setting" [ ➙ 46].
'Next calibration check' counts down from twelve to zero and will be
highlighted if zero is reached. The laser remains fully functional. Unit is
month.

Time to service

The safety test is a mandatory test for all medical devices. The
SIROLaser Advance needs to be tested once every two years.
"Time to service" counts down from 24 to zero and will be highlighted if
zero is reached. The laser remains fully functional. Unit is month.
NOTICE
Legal regulations require a regular safety test of the performance of a
laser device. The SIROLaser Advance needs to be tested once every
two years. Please contact your local dental depot or an authorized
Customer Service Department for technical support.
If internal or external calibration check failed after last calibration a
warning screen pops up after self test. The laser remains fully functional.
Operation
Self Test
37