Energy; Average Power; Back Button; Favorite Button - Sirona SiroLaser advance Operating Instructions Manual

Hide thumbs Also See for SiroLaser advance:
Table of Contents

Advertisement

 Operation
Main home screen
empty
empty
empty
empty
empty
empty
44
NOTICE
Between 1 kHz -10 kHz the duty cycle cannot be below 10% and above
90% due to physical limits. So at 1 kHz -10 kHz any input below 10% will
result in an error message. Values below 10% have to be increased to
at least 10% and values above 90% have to be decreased to at most
90%.
Also due to physical limits the chopped mode is not possible with an
average power below 0.05 W. If the user wants to set the chopped mode
below this value an error message appears: "Chopped mode available
only with an average power above 0.05 W."
The peak-pulse mode is not possible with an average power below
0.3 W. If the user wants to set the peak-pulse mode below this value an
error message appears: "Peak Pulse available only with an average
power above 0.3 W."
When returning from peak-pulse mode to chopped mode note that the
average power will remain and the power will be calculated regarding a
preset duty cycle of 50%.

6. Energy

in our example, 0 J (joule). The system calculates the energy used during
treatment (in J) from the power values and the laser activation time. So
as soon as application is started and time running the energy will be
calculated and accumulated as long as the laser is active.

7. Average power

in our example, 3 W (watt). The system calculates the average power (in
W) from the power values and the selected duty cycle.
Further more the example here shows:

8. Back button

By pushing the back button you will jump back one screen.

9. Favorite button

By pushing the 'save as favorite' button this program will be linked to the
favorite submenu.
This is the 'delete favorite' button to delete the link from the favorite
submenu.
10. Help button
By pushing the 'help' button the help menu will be opened and you can
read additional information about this treatment.
11. Laser button
By pushing the 'Activate Laser' button the laser will be made ready for
operation.

Sirona Dental Systems GmbH

Operating Instructions SIROLaser Advance
62 60 504 D3545
D3545.201.01.08.02 08.2012

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents