Figure 4-2. Manual Door Release - Beckman Coulter Avanti J-20 XP Instruction Manual

Expanded-performance centrifuge
Hide thumbs Also See for Avanti J-20 XP:
Table of Contents

Advertisement

Action
5. Pull the interlock lever out
and to the left (at about a
45-degree angle), and while
holding it out, step on the
foot pedal.
6. After the door opens, first
release the foot pedal and
then release the interlock
lever.
Inner
Front
Panel
Retaining
Strip

Figure 4-2. Manual Door Release

Result
Depending on the level of vacuum in the chamber at the time of the
power failure, the door may or may not open. If it opens, go to step 6.
If it does not open, you will need to manually vent the vacuum. Go to
step 7.
Sample can be removed. Go to step 10.
Troubleshooting
Manual Door
Release
Interlock Lever
4-7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents