Single And Three-Phase Power Connections - Beckman Coulter Avanti J-20 XP Instruction Manual

Expanded-performance centrifuge
Hide thumbs Also See for Avanti J-20 XP:
Table of Contents

Advertisement

Table 1-1. Nominal Supply Voltage Ratings for the Avanti J-20XP
Nominal Instrument
Voltage Rating
Single-phase, 200/208/240 V
Single-phase, 230 V
Three-phase, 220/380 V
+ Neutral*
* Unbalanced three-phase. Split for single-phase operation internally.

SINGLE AND THREE-PHASE POWER CONNECTIONS

!
WARNING
To reduce the risk of electrical shock, this
equipment uses a three-wire or five-wire elec-
trical cord and plug to connect the centrifuge
to earth-ground. To preserve this safety
feature, make sure that the matching wall
outlet receptacle is properly wired and
earth-grounded.
Prior to instrument purchase, the power configuration should be
determined and the appropriate instrument ordered.
Instrument
Instrument
Part Number
Part
(Elutriation
Number
System)
368608
368640
368611
368641
368612
368643
Figure 1-1 shows the power connection for single-phase centrifuges,
including earth-ground and two power leads with 30-ampere circuit
breakers. Figure 1-2 shows the power connection for three-phase
centrifuges, including earth-ground and three power leads with
minimum 16-ampere circuit breakers and common neutral
connection. Table 1-2 contains wiring information.
Specifications and Preinstallation Requirements
Nominal
Power Cord and Plug
Supply Frequency
180–264 V,
permanently attached three-
50/60 Hz, 30 A
wire UL/CSA-approved cord
with NEMA type 6-30P plug
180–264 V,
permanently attached three-
50 Hz, 30 A
wire CENELEC harmonized
cord without plug
313–457 V,
permanently attached five-
plus neutral,
wire CENELEC harmonized
50 Hz, 16 A
cord without plug
Description
1-9