Manual Door Release - Beckman Coulter Avanti J-26S XPI Instructions For Use Manual

High-performance centrifuge
Table of Contents

Advertisement

Troubleshooting
Accessing the Rotor in Case of Power Failure
4
Lift the front panel up and off the centrifuge; set it aside.
• You will see an inner front panel that extends about halfway down from the centrifuge top
(see
• Below this inner panel is a black manual door release interlock lever (see
Figure 3.2 Manual Door Release
1
2
1. Inner Front Panel
2. Retaining Strip
5
Pull the interlock lever out and to the left (at about a 45-degree angle), and while holding it out,
step on the foot pedal.
Depending on the level of vacuum in the chamber at the time of the power failure, the door may
or may not open.
a. If it opens, go to step 6.
b. If it does not open, you will need to manually vent the vacuum. Go to step 7.
6
After the door opens, first release the foot pedal and then release the interlock lever.
Sample can be removed.
a. Go to step 10.
7
To vent the chamber vacuum, grasp the red rubber hose and pull it up until it comes off of the
pump fitting (see
3-6
Figure
3.2).

3. Manual Door Release

Interlock Lever
Figure
3.3).
Figure
3
PN B10093AE
3.2).