Troubleshooting
Accessing the Rotor in Case of Power Failure
1
Turn the power switch to off (
the Mains (power) plug from the outlet receptacle.
• The Mains (power) plug is the disconnect device and must remain easily accessible.
— Position the centrifuge so that it is easy to remove the Mains (power) plug from the
2
Insert a 4-mm or
the wrench to the left (counterclockwise) until the latch disengages.
3
Repeat step 2, inserting the Allen wrench through the other hole.
• After the second latch disengages, the front panel will fall forward about an inch from
the top.
• The bottom of the front panel is held in place by three grooved brackets, attached to the
front panel, that seat over a metal retaining strip on the centrifuge bottom panel.
4
Lift the front panel up and off the centrifuge; set it aside.
• You will see an inner front panel that extends about halfway down from the centrifuge top
(see
• Below this inner panel is a black manual door release interlock lever (see
Figure 3.2 Manual Door Release
1
2
1. Inner Front Panel
3-6
) and unplug the power cord from the main source by removing
O
outlet receptacle.
5
/
-in. Allen wrench straight through one of the holes (either one) and turn
32
Figure
3.2).
2. Retaining Strip
3
3. Manual Door Release Interlock Lever
Figure
3.2).
PN B10087AB