Specifications and Preinstallation Requirements
Rotor Profile and Description
JA-20 Fixed Angle, 34° (8 place)
JA-18.1 Fixed Angle (24 place)
JA-18 Fixed Angle, 23° (10 place)
JA-17 Fixed Angle, 25° (6 place)
JLA-16.250 Fixed Angle, 25° (6 place)
JA-14 Fixed Angle, 25° (6 place)
a
Maximum speeds are based on a solution density of 1.2 g/mL with the following exceptions: the JA-18.1 rotor is rated for a density of 1.4 g/mL;
the JCF-Z rotor is rated for a density of 1.45 g/mL; and the JE-5.0 rotor is rated for a density of 3 g/mL.
b
Maximum speed for the JA-30.50 Ti rotor is 26 000 rpm in the Avanti J-20XP centrifuge, and 30 000 rpm in the Avanti J-30I centrifuge.
c
Maximum speed for the JS-24.38 and JS-24.15 rotors is 10 000 rpm in the Avanti J-20XP centrifuge,
and 24 000 rpm in the Avanti J-30I centrifuge.
d
In the Avanti J-20XP, the JCF-Z Continuous Flow/Zonal Rotor can be used in the continuous flow and reorienting gradient zonal
configurations only.
1-4
Max
a
RPM
r
= 108 mm
20 000
max
45° adapter
18 000
r
= 116 mm
max
25° adapter
17 000
r
= 112 mm
max
18 000
r
= 132 mm
max
r
= 132 mm
17 000
max
r
= 134 mm
16 000
max
r
= 137 mm
14 000
max
Max RCF
Max
( × g )
Capacity
8 × 50 mL
48 400
24 × 1.8 mL
42 100
24 × 1.8 mL
36 300
10 × 100 mL
47 900
14 × 50 mL
39 800
6 × 250 mL
38 400
6 × 250 mL
30 100
Continued —
Rotor
Manual
Number
J-TB-003
J-TB-037
J-TB-035
J-TB-017
J-TB-072
J-TB-004