Disabling The Audible Alarms; Table 2. Alarm Limit, Breath Beep Pitch, Calendar, And Clock Mode Parameter - Nonin 9847V Operator's Manual

Veterinary handheld pulse; oximeter and carbon dioxide (co2) detector
Table of Contents

Advertisement

Table 2: Alarm Limit, Breath Beep Pitch, Calendar, and Clock Mode Parameter
Parameter
Alarm Mode
dFt =default limits
rCL = recall stored limits
C02 = only no breath and the sensor
alarms
No Breath Delay (seconds)
Oxygen Saturation Low Limit (%)
Pulse Rate Low Limit (beats/minute)
Pulse Rate High Limit (beats/minute)
Oxygen Saturation High Limit (%)
Breath Beep Pitch (fixed or variable)
Year
Month
Day
Hours
Minutes
a.The default setting for 02H is 0FF (disabled). Any of the other animal alarm limit parameters can be individually
disabled by setting to 0FF.

Disabling the Audible Alarms

The audible alarms can be temporarily disabled, or they can be permanently disabled at power on
when entering setup mode using the Audible Alarm Disable button.
When the audible alarms are disabled, informational beeps continue to sound at the set volume, and
the visible alarm indicators remain enabled.
Temporarily Disabling the Audible Alarms (Maximum 2 Minutes)
1. With the unit on, press and then release Audible Alarm Disable.
2. The Audible Alarm Disabled indicator blinks.
3. The audible alarms will be disabled for 2 minutes (unless overridden by an audible critical
battery alarm).
4. The audible alarms can be re-enabled by pressing Audible Alarm Disable again before the
2-minute period has expired.
20
Appears in
Pulse Rate Display
SpO
Display
Range of Values
2
ALr
dFt, rCL, C02
nbd
15 to 60 by 5, OFF
02L
50 to 95 by 1, OFF
H L
20 to 200 by 5, OFF
H H
100 to 425 by 25, OFF,
50 to 100 by 5
a
02H
80 to 100 by 1, OFF
FPt
ON or OFF
y
00 - 99
nn
00 - 12
d
01 - 31
h
00 - 23
nn
00 - 59
Using the 9847V Pulse Oximeter
Default Value
dFt
20
85
50
200
OFF
OFF
(fixed pitch off)
09
00
00
00
00