2. Release the audible alarm disable button when "888 888" is displayed on the %SpO
pulse rate displays. After the power on self-test, the "audible alarm disabled" indicator will
illuminate steadily. The patient alarms will be set at the default limits. The audible alarms can be
re-enabled by pressing the audible alarm disable button again.

Table 2: Patient Alarm Limit, Breath Beep Pitch, Calendar, and Clock Mode Parameters

Parameter
Alarm Mode
(dFt =default limits;
rCL = recall stored limits;
C02 = only no breath and the sensor
alarms)
No Breath Delay (seconds)
Oxygen Saturation Low Limit (%)
Pulse Rate Low Limit (beats/minute)
Pulse Rate High Limit (beats/minute)
Oxygen Saturation High Limit (%)
Breath Beep Pitch (fixed or variable)
Year
Month
Day
Hours
Minutes
a.The default setting for "02H" is "OFF" (disabled). Any of the other patient alarm limit parameters can be individually
disabled by setting to "OFF".
18
Appears in
Pulse Rate ♥ Display
SpO
Display
Range of Values
2
ALr
dFt, rCL, C02
nbd
15 to 60 by 5, OFF
02L
50 to 95 by 1, OFF
H L
30 to 110 by 5, OFF
H H
75 to 275 by 5, OFF
a
02H
80 to 100 by 1, OFF
FPt
ON or OFF
y
00
99
-
nn
00
12
-
d
01
31
-
h
00
23
-
nn
00
59
-
Using the 9847 Pulse Oximeter
and
2
Default Value
dFt
20
85
50
200
OFF
OFF (fixed pitch off)
09
00
00
00
00