Alarm Summary
If patient SpO
or pulse readings are equal to or above the upper alarm limit, or if
2
they are equal to or below the lower alarm limit, the display unit will signal an alarm.
High Priority Alarm Description
SpO
Upper Alarm Limit
2
SpO
Lower Alarm Limit
2
Pulse Upper Alarm Limit
Pulse Lower Alarm Limit
Silencing Alarms
Momentary (2-Minute) Alarm Silence
To silence alarms for two minutes, press the Alarm Silence button.
Continuous Alarm Silence
In order to permanently silence all alarms, DIP switch 2 must be placed in the UP
position. This allows the alarm volume to be set to zero. The Alarm Silence LED will
remain illuminated when the alarm volume is set to zero. Refer to "DIP Switches" for more
information.
Factory
Default
Adjustment Range
Off
Off, 80 to 100
80%
Off, 50 to 95
200 BPM
Off, 75 to 275
50 BPM
Off, 30 to 110
Step Value
1% SpO
2
1% SpO
2
5 BPM
5 BPM
27