Connecting The Gp-Ib Cable - YOKOGAWA wt1600 User Manual

Digital power meter
Hide thumbs Also See for wt1600:
Table of Contents

Advertisement

1.2

Connecting the GP-IB Cable

GP-IB Cable
Connection Method
Connection Precautions
1-2
The GP-IB connector on the side panel of the PZ4000 is a 24-pin connector that
conforms to IEEE Standard 488-1978. Use a GP-IB cable that also conforms to IEEE
Standard 488-1978.
Connect the GP-IB cable as shown below.
• Be sure to tighten the screws on the GP-IB cable connector firmly.
• The instrument can be connected to more than one item of equipment (e.g. a personal
computer) if more than one GP-IB cable is used. However, it is not possible to
connect more than 15 items of equipment (including the controller) to a single bus.
• If you connect the instrument to more than one item of equipment, make sure that a
different address is used for each item.
• Each connecting cable must be 2 m or less in length.
• The total length of all the cables must not exceed 20 m.
• While communications are in progress, more than two-thirds of the connected
equipment items must be turned ON.
• When connecting more than one item of equipment, connect them so that the
connection route forms a star or linear configuration. Loop or parallel wiring is not
allowed.
Be sure to switch off power to both your PC and the oscilloscope before
connecting or disconnecting cables. Failure to switch power off may
cause internal circuit failure or improper operation.