:STORe:FILE:COMMent
Function
Sets the comment to be added to the file when
saving the stored data or queries the current
setting.
Syntax
:STORe:FILE:COMMent {<String>}
:STORe:FILE:COMMent?
<string> = 25 characters or less
Example
:STORE:FILE:COMMENT "CASE1"
:STORE:FILE:COMMENT? ->
:STORE:FILE:COMMENT "CASE1"
:STORe:FILE:NAME
Function
Sets the name of the file when saving the stored
data or queries the current setting.
Syntax
:STORe:FILE:NAME {<Filename>}
:STORe:FILE:NAME?
Example
:STORE:FILE:NAME "DATA1"
:STORE:FILE:NAME? -> :STORE:FILE:
NAME "DATA1"
Description The save destination of the stored data is
specified using:
• the ":FILE:DRIVe" command for the drive.
• the ":FILE:CDIRectory" command for the
directory.
The save destination path can be queried
using the ":FILE:PATH?" command.
:STORe:INTerval
Function
Sets the store interval or queries the current
setting.
Syntax
:STORe:INTerval {<NRf>,<NRf>,<NRf>}
:STORe:INTerval?
1st <NRf> = 0 to 99 (hour)
2nd <NRf> = 0 to 59 (minute)
3rd <NRf> = 1 to 59 (second)
Example
:STORE:INTERVAL 0,0,0
:STORE:INTERVAL? -> :STORE:
INTERVAL 0,0,0
:STORe:ITEM
Function
Sets the items to be stored or queries the
current setting.
Syntax
:STORe:ITEM {NUMeric|WAVE|NWAVe}
:STORe:ITEM?
NUMeric = Store only the numerical values.
WAVE = Store only the waveforms.
NWAVe = Store both the numerical values and
the waveforms.
Example
:STORE:ITEM NUMERIC
:STORE:ITEM? -> :STORE:ITEM NUMERIC
:STORe:MEMory:CONVert:ABORt
Function
Abort converting the stored data from the
memory to the file.
Syntax
:STORe:MEMory:CONVert:ABORt
Example
:STORE:MEMORY:CONVERT:ABORT
IM 760101-11E
:STORe:MEMory:CONVert:EXECute
Function
Executes the converting of the stored data from
the memory to the file.
Syntax
:STORe:MEMory:CONVert:EXECute
Example
:STORE:MEMORY:CONVERT:EXECUTE
Description • The convert destination file is set using the
":STORe:FILE:..." command.
• When executing file conversion, the
instrument accesses the file twice.
When checking the completion status of the
file conversion, you must use the
"COMMUNICATE: WAIT 64" command
(checks for changes in bit 6 (ACS) of the
status register) two times as in the following
example.
"STATUS:EESR?"
Clears the extended event register.
"STORE:MEMORY:CONVERT:EXECUTE"
Starts file conversion.
"COMMUNICATE:WAIT 64"
Waits for conversion to complete, 1st time.
"STATUS:EESR?"
Clears the extended event register.
"COMMUNICATE:WAIT 64"
Waits for conversion to complete, 2nd time.
"STATUS:EESR?"
Clears the extended event register.
: S T O R e : M E M o r y : I N I T i a l i z e
Function
Executes the initialization of the storage
memory.
Syntax
:STORe:MEMory:INITialize
Example
:STORE:MEMORY:INITIALIZE
:STORe:MODE
Function
Sets the data storage/recall or queries the
current setting.
Syntax
:STORe:MODE {STORe|RECall}
:STORe:MODE?
Example
:STORE:MODE STORE
:STORE:MODE? -> :STORE:MODE STORE
:STORe:NUMeric?
Function
Queries all settings related to the storage of
numerical data.
Syntax
:STORe:NUMeric?
Example
:STORE:NUMERIC? -> :STORE:NUMERIC:
NORMAL:ELEMENT1 1;ELEMENT2 0;
ELEMENT3 0;ELEMENT4 0;ELEMENT5 0;
ELEMENT6 0;SIGMA 0;SIGMB 0;SIGMC 0;
URMS 1;UMN 1;UDC 1;UAC 1;IRMS 1;
IMN 1;IDC 1;IAC 1;P 1;S 1;Q 1;
LAMBDA 1;PHI 1;FU 1;FI 1;UPPEAK 1;
UMPEAK 1;IPPEAK 1;IMPEAK 1;CFU 1;
CFI 1;FFU 1;FFI 1;Z 1;RS 1;XS 1;
RP 1;XP 1;PC 1;TIME 0;WH 0;WHP 0;
WHM 0;AH 0;AHP 0;AHM 0;ETA 0;
SETA 0;F1 0;F2 0;F3 0;F4 0;DURMS 0;
DUMN 0;DUDC 0;DUAC 0;DIRMS 0;
DIMN 0;DIDC 0;DIAC 0