YOKOGAWA wt1600 User Manual page 108

Digital power meter
Hide thumbs Also See for wt1600:
Table of Contents

Advertisement

5.17 STATus Group
:STATus?
Function
Queries all settings related to the
communication status function.
Syntax
:STATus?
Example
:STATUS? -> :STATUS:EESE 0;FILTER1
NEVER;FILTER2 NEVER;FILTER3 NEVER;
FILTER4 NEVER;FILTER5 NEVER;
FILTER6 NEVER;FILTER7 NEVER;
FILTER8 NEVER;FILTER9 NEVER;
FILTER10 NEVER;FILTER11 NEVER;
FILTER12 NEVER;FILTER13 NEVER;
FILTER14 NEVER;FILTER15 NEVER;
FILTER16 NEVER;QENABLE 0;QMESSAGE 1
:STATus:CONDition?
Function
Queries the contents of the condition register.
Syntax
:STATus:CONDition?
Example
:STATUS:CONDITION? -> 16
Description For the description regarding how to
synchronize the program using :STATus:
CONDition, see page 4-8.
:STATus:EESE
(Extended
Event
Function
Sets the extended event enable register or
queries the current setting.
Syntax
:STATus:EESE <Register>
:STATus:EESE?
<Register> = 0 to 65535
Example
:STATUS:EESE #B0000000000000000
:STATUS:EESE? -> :STATUS:EESE 0
:STATus:EESR?
(Extended
Event
Function
Queries the content of the extended event
register and clears the register.
Syntax
:STATus:EESR?
Example
:STATUS:EESR? -> 0
:STATus:ERRor?
Function
Queries the error code and message
information (top of the error queue).
Syntax
:STATus:ERRor?
Example
:STATUS:ERROR? ->
113,"Underfined Header"
Description • When there is no error, "0, "No error"" is
returned.
• The message cannot be returned in
Japanese.
• You can specify whether or not to add the
message using the "STATus:QMESsage"
command.
5-74
Status
Enable
Status
Register)
:STATus:FILTer<x>
Function
Sets the transition filter or queries the current
setting.
Syntax
:STATus:FILTer<x> {RISE|FALL|BOTH|
NEVer}
:STATus:FILTer<x>?
<x> = 1 to 16
Example
:STATUS:FILTER2 RISE
:STATUS:FILTER2? -> :STATUS:
FILTER2 RISE
Description Specify how each bit of the condition register is
to change to set the event. If "RISE" is specified,
the event is set when the bit changes from "0" to
"1."
:STATus:QENable
Function
Sets whether or not to store messages other
than errors to the error queue (ON/OFF) or
queries the current setting.
Syntax
:STATus:QENable {<Boolean>}
:STATus:QENable?
Example
:STATUS:QENABLE ON
:STATUS:QENABLE? -> :STATUS:
QENABLE 1
register)
:STATus:QMESsage
Function
Sets whether or not to attach message
information to the response to the
"STATus:ERRor?" query (ON/OFF) or queries
the current setting.
Syntax
:STATus:QMESsage {<Boolean>}
:STATus:QMESsage?
Example
:STATUS:QMESSAGE ON
:STATUS:QMESSAGE? -> :STATUS:
QMESSAGE 1
:STATus:SPOLI?
Function
Executes serial polling.
Syntax
:STATus:SPOLl?
Example
:STATUS:SPOLL? -> :STATUS:SPOLL 0
Description This is a command specific to the serial (RS-
232) interface. An interface message is
available for the GP-IB interface.
(Serial
Poll)
IM 760101-11E

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents