YOKOGAWA wt1600 User Manual page 73

Digital power meter
Hide thumbs Also See for wt1600:
Table of Contents

Advertisement

:HARMonics:ORDer
Function
Sets the maximum and minimum orders to be
analyzed or queries the current setting.
Syntax
:HARMonics:ORDer {<NRf>,<NRf>}
:HARMonics:ORDer?
1st <NRf> = 0 or 1 (minimum order to be
analyzed)
2nd <NRf> = 1 to 100 (maximum order to be
analyzed)
Example
:HARMONICS:ORDER 1,100
:HARMONICS:ORDER? -> :HARMONICS:
ORDER 1,100
:HARMonics:PLLSource
Function
Sets the PLL source or queries the current
setting.
Syntax
:HARMonics:PLLSource {U<x>|I<x>|
EXTernal|SAMPle}
:HARMonics:PLLSource?
<x> = 1 to 6 (element)
EXTernal = External clock input (Ext Clk)
SAMPle = Sampling clock input (Smp Clk)
Example
:HARMONICS:PLLSOURCE U1
:HARMONICS:PLLSOURCE? ->:HARMONICS:
PLLSOURCE U1
:HARMonics[:STATe]
Function
Turns ON/OFF the harmonic measurement
mode or queries the current setting.
Syntax
:HARMonics[:STATe] {<Boolean>}
:HARMonics:STATe?
Example
:HARMONICS:STATE ON
:HARMONICS:STATE? -> :HARMONICS:
STATE 1
:HARMonics:THD
Function
Sets the equation used to calculate the THD
(total harmonic distortion) or queries the current
setting.
Syntax
:HARMonics:THD {TOTal|FUNDamental}
:HARMonics:THD?
Example
:HARMONICS:THD TOTAL
:HARMONICS:THD? -> :HARMONICS:
THD TOTAL
:HARMonics:WIDTh
Function
Sets the data length used during harmonic
measurement or queries the current setting.
Syntax
:HARMonics:WIDTh {<NRf>}
:HARMonics:WIDTh?
<NRf> = 2048, 4096, 8192
Example
:HARMONICS:WIDTH 8192
:HARMONICS:WIDTH? -> :HARMONICS:
WIDTH 8192
IM 760101-11E
5.7 HARMonics Group
5-39
5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents