3.3 Connecting the WT to a PC
Explanation
3-4
14.
Turn the jog shuttle to set the timeout time.
For instructions on keyboard operations, see section 3.12 "Entering Values and Strings" in
user's manual IM760101-01E.
15.
Press SELECT or ESC to close the box.
Entering TCP/IP Settings
You must enter TCP/IP settings to control the WT from a PC using the Ethernet
interface. For instructions on entering settings, see section 13.2 "Setting the Ethernet
Interface (TCP/IP)" in the user's manual IM760101-01E.
You can control the WT from a PC using the Ethernet interface. To enable this function, you
must confirm that your WT is running ROM version 2.01 or later, and that YOKOGAWA's
dedicated software has been installed on the PC according to the instructions above.
Retail Software
WTViewer (Model 760122) version 2.00 or later.
A trial version is available and can be downloaded from the following URL.
http://www.yokogawa.com/tm/760122/
Free Software
Wirepuller version 1.02 or later.
Wirepuller can be downloaded from the following URL.
http://www.yokogawa.com/tm/wirepuller/
Setting the User Name
• Enter the user name to allow access to the WT1600.
• Enter up to 15 characters.
• The characters that can be used are 0-9, A-Z, %, _, ( ) (parenthesis), - (minus sign).
• If you specify anonymous, the WT1600 can be accessed from the outside (PC)
without a password.
Setting the Password
• Enter the password for the user name to allow access to the WT1600.
• Enter up to 15 characters.
• The characters that can be used are 0-9, A-Z, %, _, ( ) (parenthesis), - (minus sign).
• If the user name is set to anonymous, the WT1600 can be accessed from the outside
(PC) without a password. The password setting is entered twice.
Setting the Timeout Time
The WT1600 closes the connection to the network if there is no access for a certain
period of time (timeout time).
The available settings are 0 to 3600 s, or Infinite. The default value is Infinite.
Note
The settings will not take effect until the unit is turned OFF then back ON again.
ESC
IM 760101-11E