YOKOGAWA wt1600 User Manual page 93

Digital power meter
Hide thumbs Also See for wt1600:
Table of Contents

Advertisement

SYNChroniz e
<Space>
?
MHOLd
<Space>
?
?
:MEASure?
Function
Queries all settings related to the measurement.
Syntax
:MEASure?
Example
• Example for normal measurement
:MEASURE? -> :MEASURE:AVERAGING:
STATE 0;TYPE EXPONENT;COUNT 2;:
MEASURE:FREQUENCY:ITEM U1,I1,U2;:
MEASURE:FUNCTION1:STATE 0;
EXPRESSION "URMS(E1)";UNIT "V";:
MEASURE:FUNCTION2:STATE 0;
EXPRESSION "IRMS(E1)";UNIT "A";:
MEASURE:FUNCTION3:STATE 0;
EXPRESSION "UPPK(E1)";UNIT "V";:
MEASURE:FUNCTION4:STATE 0;
EXPRESSION "IPPK(E1)";UNIT "A";:
MEASURE:DMEASURE:OBJECT SIGMA;
TYPE OFF;:MEASURE:SFORMULA RMS;
PC:IEC 1976;P1 0.5000;P2 0.5000;:
MEASURE:PHASE 180;
SYNCHRONIZE MASTER;MHOLD 0
• Example for harmonic measurement
:MEASURE? -> :MEASURE:AVERAGING:
STATE 0;:MEASURE:FREQUENCY:
ITEM U1,I1,U2;:MEASURE:FUNCTION1:
STATE 0;EXPRESSION "U(E1,OR1)";
UNIT "V";:MEASURE:FUNCTION2:
STATE 0;EXPRESSION "I(E1,OR1)";
UNIT "A";:MEASURE:FUNCTION3:
STATE 0;EXPRESSION "P(E1,OR1)";
UNIT "W";:MEASURE:FUNCTION4:
STATE 0;EXPRESSION "S(E1,OR1)";
UNIT "VA";:MEASURE:
SYNCHRONIZE MASTER
:MEASure:AVERaging?
Function
Queries all settings related to averaging.
Syntax
:MEASure:AVERaging?
Example
• Example for normal measurement
:MEASURE:AVERAGING? -> :MEASURE:
AVERAGING:STATE 1;TYPE EXPONENT;
COUNT 2
• Example for harmonic measurement
:MEASURE:AVERAGING? -> :MEASURE:
AVERAGING:STATE 1
IM 760101-11E
MASTer
SLAVe
OFF
ON
<NRf>
:MEASure:AVERaging:COUNt
Function
Sets the averaging coefficient or queries the
current setting.
Syntax
:MEASure:AVERaging:COUNt {<NRf>}
:MEASure:AVERaging:COUNt?
<NRf> = 2, 4, 8, 16, 32, 64 (when TYPE =
EXPonent)
<NRf> = 8, 16, 32, 64, 128, 256 (when TYPE =
LINear)
Example
:MEASURE:AVERAGING:COUNT 2
:MEASURE:AVERAGING:COUNT? ->
:MEASURE:AVERAGING:COUNT 2
Description This command is valid only during normal
measurement. For details on the averaging
coefficient (attenuation constant) during
harmonic measurement, see the WT1600
User's Manual.
:MEASure:AVERaging[:STATe]
Function
Turns ON/OFF averaging or queries the current
setting.
Syntax
:MEASure:AVERaging
[:STATe] {<Boolean>}
:MEASure:AVERaging:STATe?
Example
:MEASURE:AVERAGING:STATE ON
:MEASURE:AVERAGING:STATE? ->
:MEASURE:AVERAGING:STATE 1
Description The averaging for harmonic measurement can
only be turned ON/OFF. For details on the
averaging during harmonic measurement, see
the WT1600 User's Manual.
:MEASure:AVERaging:TYPE
Function
Sets the averaging type or queries the current
setting.
Syntax
:MEASure:AVERaging:TYPE {EXPonent|
LINear}
:MEASure:AVERaging:TYPE?
Example
:MEASURE:AVERAGING:TYPE EXPONENT
:MEASURE:AVERAGING:TYPE? ->
:MEASURE:AVERAGING:TYPE EXPONENT
Description This command is valid only during normal
measurement. For details on the averaging type
during harmonic measurement, see the
WT1600 User's Manual.
5.13 MEASure Group