5.11 INPut Group
• When the crest factor is set to 6
• For a 5-A input element
<Current> = 5, 10, 25, 50, 100, 250 (mA),
0.5, 1, 2.5 (A) (when TERMinal = DIRect)
<Current> = 25, 50, 125, 250 (mV), 0.5,
1.25, 2.5, 5 (V) (when TERMinal = SENSor)
AUTO = Auto range
• For a 50-A input element
<Current> = 0.5, 1, 2.5, 5, 10, 25 (A)
(when TERMinal = DIRect)
<Voltage > = 25, 50, 125, 250 (mV), 0.5,
1.25, 2.5, 5 (V) (when TERMinal SENSor)
AUTO = Auto range
Example
:INPUT:CURRENT:RANGE:ELEMENT1 5A
:INPUT:CURRENT:RANGE:ELEMENT1? ->
:INPUT:CURRENT:RANGE:
ELEMENT1 5.00E+00
Description • The selectable range is determined by the
input element type (5A/50A) of the target
element and the current measurement
terminal setting ([:INPut]:CURRent:
TERMinal:ELEMent<x>).
• Specifying "AUTO" with this command is
equivalent to setting "[:INPut]:CURRent:
AUTO:ELEMent<x>" to "ON."
[:INPut]:CURRent:SRATio?
Function
Queries the current sensor scaling constants of
all elements.
Syntax
[:INPut]:CURRent:SRATio?
Example
:INPUT:CURRENT:SRATIO? -> :INPUT:
CURRENT:SRATIO:ELEMENT1 10.0000;
ELEMENT2 10.0000;ELEMENT3 10.0000;
ELEMENT4 10.0000;ELEMENT5 10.0000;
ELEMENT6 10.0000
[ : I N P u t ] : C U R R e n t : S R A T i o [ : A L L ]
Function
Collectively sets the current sensor scaling
constants of all elements.
Syntax
[:INPut]:CURRent:SRATio
[:ALL] {<NRf>}
<NRf> = 0.0001 to 99999.9999
Example
:INPUT:CURRENT:SRATIO:ALL 10
[:INPut]:CURRent:SRATio:ELEMent<x>
Function
Sets the current sensor scaling constant of the
element or queries the current setting.
Syntax
[:INPut]:CURRent:SRATio:
ELEMent<x> {<NRf>}
[:INPut]:CURRent:SRATio:ELEMent<x>?
<x> = 1 to 6
<NRf> = 0.0001 to 99999.9999
Example
:INPUT:CURRENT:SRATIO:ELEMENT1 10
:INPUT:CURRENT:SRATIO:ELEMENT1? ->
:INPUT:CURRENT:SRATIO:
ELEMENT1 10.0000
5-50
[:INPut]:CURRent:TERMinal?
Function
Queries the current measurement terminal of all
elements.
Syntax
[:INPut]:CURRent:TERMinal?
Example
:INPUT:CURRENT:TERMINAL? -> :INPUT:
CURRENT:TERMINAL:ELEMENT1 DIRECT;
ELEMENT2 DIRECT;ELEMENT3 DIRECT;
ELEMENT4 DIRECT;ELEMENT5 DIRECT;
ELEMENT6 DIRECT
[ : I N P u t ] : C U R R e n t : T E R M i n a l [ : A L L ]
Function
Collectively sets the current measurement
terminals of all elements.
Syntax
[:INPut]:CURRent:TERMinal
[:ALL] {DIRect|SENSor}
DIRect = Direct input
SENSor = Current sensor input
Example
:INPUT:CURRENT:TERMINAL:ALL DIRECT
[:INPut]:CURRent:TERMinal:ELEMent<x>
Function
Sets the current measurement terminal of the
element or queries the current setting.
Syntax
[:INPut]:CURRent:TERMinal:
ELEMent<x> {DIRect|SENSor}
[:INPut]:CURRent:TERMinal:
ELEMent<x>?
<x> = 1 to 6
DIRect = Direct input
SENSor = Current sensor input
Example
:INPUT:CURRENT:TERMINAL:
ELEMENT1 DIRECT
:INPUT:CURRENT:TERMINAL:ELEMENT1?
-> :INPUT:CURRENT:TERMINAL:
ELEMENT1 DIRECT
[ : I N P u t ] : F I L T e r ?
Function
Queries all settings related to the filter.
Syntax
[:INPut]:FILTer?
Example
:INPUT:FILTER? -> :INPUT:FILTER:
LINE:ELEMENT1 OFF;ELEMENT2 OFF;
ELEMENT3 OFF;ELEMENT4 OFF;
ELEMENT5 OFF;ELEMENT6 OFF;:INPUT:
FILTER:ZCROSS:ELEMENT1 OFF;
ELEMENT2 OFF;ELEMENT3 OFF;
ELEMENT4 OFF;ELEMENT5 OFF;
ELEMENT6 OFF
[ : I N P u t ] : F I L T e r : L I N E ?
Function
Queries the line filter settings of all elements.
Syntax
[:INPut]:FILTer:LINE?
Example
:INPUT:FILTER:LINE? -> :INPUT:
FILTER:LINE:ELEMENT1 OFF;
ELEMENT2 OFF;ELEMENT3 OFF;
ELEMENT4 OFF;ELEMENT5 OFF;
ELEMENT6 OFF