:MEASure:FUNCtion<x>:UNIT
Function
Sets the unit to be added to the computation
result of the user-defined function or queries the
current setting.
Syntax
:MEASure:FUNCtion<x>:
UNIT {<String>}
:MEASure:FUNCtion<x>:UNIT?
<x> = 1 to 4
<string> = 8 characters or less
Example
:MEASURE:FUNCTION1:UNIT "V"
:MEASURE:FUNCTION1:UNIT? ->
:MEASURE:FUNCTION1:UNIT "V"
Description • Only the characters and symbols displayed
on the keyboard on the screen can be used.
• This command does not affect the
computation result.
:MEASure:MHOLd
Function
Turns ON/OFF the MAX HOLD function or
queries the current setting.
Syntax
:MEASure:MHOLd {<Boolean>}
:MEASure:MHOLd?
Example
:MEASURE:MHOLD ON
:MEASURE:MHOLD? -> :MEASURE:MHOLD 1
Description This command is valid only during normal
measurement.
:MEASure:PC?
Function
Queries all settings related to the calculation of
Pc (Corrected Power).
Syntax
:MEASure:PC?
Example
:MEASURE:PC? -> :MEASURE:PC:
IEC 1976;P1 0.5000;P2 0.5000
:MEASure:PC:IEC
Function
Sets the equation used to calculate Pc
(Corrected Power) or queries the current
setting.
Syntax
:MEASure:PC:IEC {<NRf>}
:MEASure:PC:IEC?
<NRf> = 1976, 1993
Example
:MEASURE:PC:IEC 1976
:MEASURE:PC:IEC? -> :MEASURE:PC:
IEC 1976
Description • This command is valid only during normal
measurement.
• Specify the year when the equation used to
calculate the Pc was issued by IEC76-1.
IM 760101-11E
:MEASure:PC:P<x>
Function
Sets the parameter used to calculate Pc
(Corrected Power) or queries the current
setting.
Syntax
:MEASure:PC:P<x> {<NRf>}
:MEASure:PC:P<x>?
<x> = 1, 2
<NRf> = 0.0000 to 9.9999
Example
:MEASURE:PC:P1 0.5
:MEASURE:PC:P1? -> :MEASURE:PC:
P1 0.5000
Description • This command is valid only during normal
measurement.
• This parameter is used when the
":MEASure:PC:IEC" setting is set to
"1976(IEC76-1(1976))."
:MEASure:PHASe
Function
Sets the display format of the phase difference
or queries the current setting.
Syntax
:MEASure:PHASe {<NRf>}
:MEASure:PHASe?
<NRf> = 180, 360
Example
:MEASURE:PHASE 180
:MEASURE:PHASE? -> :MEASURE:
PHASE 180
Description • This command is valid only during normal
measurement.
• Displays the phase using ±0 to 180° (Lead/
Lag) for "180" and 0 to 360° for "360."
:MEASure:SFORmula
Function
Sets the equation used to calculate S (reactive
power) or queries the current setting.
Syntax
:MEASure:SFORmula {RMS|MEAN|DC|
MRMS}
:MEASure:SFORmula?
Example
:MEASURE:SFORMULA RMS
:MEASURE:SFORMULA? -> :MEASURE:
SFORMULA RMS
Description • This command is valid only during normal
measurement.
• The correspondence between the selections
and equations is as follows.
RMS: S = Urms * Irms
MEAN: S = Umean * Imean
DC: S = Udc * Idc
MRMS: S = Umean * Irms
:MEASure:SYNChronize
Function
Sets the synchronized measurement mode or
queries the current setting.
Syntax
:MEASure:SYNChronize {MASTer|SLAVe}
:MEASure:SYNChronize?
Example
:MEASURE:SYNCHRONIZE MASTER
:MEASURE:SYNCHRONIZE? -> :MEASURE:
SYNCHRONIZE MASTER
5.13 MEASure Group
5-61
5