:MOTor:TORQue:SCALing
Function
Sets the scaling factor for torque computation or
queries the current setting.
Syntax
:MOTor:TORQue:SCALing {<NRf>}
:MOTor:TORQue:SCALing?
<NRf> = 0.0001 to 99999.9999
Example
:MOTOR:TORQUE:SCALING 1
:MOTOR:TORQUE:SCALING? -> :MOTOR:
TORQUE:SCALING 1.0000
:MOTor:TORQue:UNIT
Function
Sets the unit to add to the torque computation
result or queries the current setting.
Syntax
:MOTor:TORQue:UNIT {<String>}
:MOTor:TORQue:UNIT?
<string> = 8 characters or less
Example
:MOTOR:TORQUE:UNIT "Nm"
:MOTOR:TORQUE:UNIT? -> :MOTOR:
TORQUE:UNIT "Nm"
Description • Only the characters and symbols displayed
on the keyboard on the screen can be used.
• This command does not affect the
computation result.